The RF IgA Calibrator component of the EL-RF/3 Rheumatoid Factor test system may be unstable, resulting in a drop in the RF IgA calibrator OD value which could invalidate an RF IgA test run.
Source facts
Action
TheraTest sent "Urgent Device Recall" letters dated March 28, 2012 to customers that received the affected kit, requesting them to return or destroy the remainder of the product. TheraTest issued a replacement kit to all customers that had unused product as well as any material they consumed which resulted in a failed run. Customers were also instructed to complete the enclosed confirmation form and fax the completed form back to TheraTest. Any questions were directed to TheraTest at 800-441-0771.
Initiated
2012-03-29
Root cause
Nonconforming Material/Component
Terminated
2017-09-05
Distribution
Nationwide Distribution (US)
Recall number
Z-1546-2012
Recall status
Terminated
CDRH record ID
108288
FDA classified
2012-05-14
Recall event number
61500