Open original FDA record ↗Recall record
- FDA identifier
- Z-1546-2015
- Company in FDA record
- Pega Medical Inc.
- FDA classification date
- Apr 29, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.
Source facts
- Action
- The firm sent the Urgent Field Safety Notice-Recall letter, dated April 10, 2015 to consignees via email. The letter listed the affected product as FEMALE DRIVER FASSIER-DUVAL IM TELESCOPIC SYSTEM. Consignees are requested to return devices back to the firm for disposal. Consignees with questions should contact the firm at 450-688-5144 ext. 242 or send email to egarcia@pegamedical.com
- Initiated
- 2015-04-10
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-06-29
- Distribution
- Worldwide distribution. US in Delaware, Florida, Nebraska, and New York; Australia, Brazil, and United Kingdom.
- Recall number
- Z-1546-2015
- Recall status
- Terminated
- CDRH record ID
- 135963
- FDA classified
- 2015-04-29
- Recall event number
- 71061