Open original FDA record ↗Recall record
- FDA identifier
- Z-1546-2024
- Company in FDA record
- Beckman Coulter, Inc.
- FDA classification date
- Aug 30, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Source facts
- Action
- Beckman Coulter notified consignees on 03/05/2024 via FedEx and email. The letter outlined the reason for the recall and requested consignees to review inventory for affected units. If any affected units are identified, to inspect the area underneath the wash buffer drawer for evidence of leaking wash buffer. If leaking or pooling is observed, clean up liquid wearing appropriate Personal Protective Equipment (PPE), discontinue the use of the DxI 9000, and contact your Beckman Coulter representative immediately. If no leaking or pooling of Wash Buffer in observed, continue operating the instrument as normal and continue to monitor for leaking. A Field Service Representative will contact customers to replace the internal wash buffer tank on the affected instruments. Consignees were also instructed to share this information with laboratory staff, notify any customers if further distributed, and complete and return the response form.
- Initiated
- 2024-03-05
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.
- Recall number
- Z-1546-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206540
- FDA classified
- 2024-08-30
- Recall event number
- 94191