Skip to content

Public FDA record

Z-1547-2024: Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units

Recall record Z-1547-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1547-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Sep 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury

Source facts

Action
Olympus Corporation of the Americas issued Medical Device Safety Reminder letter on 3/1/24 via UPS. Letter states reason for recall, health risk and action to take: 1. Follow your facility's procedures for communication and handling of Field Safety Notices. Ensure all personnel, including clinical staff, are informed of the contents of this letter and the Instructions for Use. You may add a copy of this letter with your IFU. 2. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall. b. Enter the recall number: 0443 c. Complete the form as instructed. 3. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. Olympus requests you to report any complaints, including any injuries associated with Triangle Tip Electrosurgical knife tip breakage, the Olympus Technical Assistance Center at 1-800-848-9024, option 1. If you require additional information, please do not hesitate to contact me directly at 647-999-3203 or Cynthia.Ow@Olympus.com.
Initiated
2024-03-01
Root cause
Use error
Distribution
Nationwide
Recall number
Z-1547-2024
Recall status
Open, Classified
CDRH record ID
206516
FDA classified
2024-09-12
Recall event number
94173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗