Open original FDA record ↗Recall record
- FDA identifier
- Z-1549-06
- Company in FDA record
- Pointe Scientific, Inc.
- FDA classification date
- Sep 30, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability.
Source facts
- Action
- Consignees were notified via recall letter dated 6/9/06 and a second letter dated 6/21/06 extending the recall to additional products/lots, to cease using the products, conduct a recall to the user level, destroy products on hand and to notify the firm via the response form of the amount destroyed for replacement purposes.
- Initiated
- 2006-06-09
- Root cause
- Other
- Terminated
- 2007-11-20
- Distribution
- Worldwide distribution ---- including USA and countries of Canada, Ecuador, Germany, Greece, India, Mexico, Peru, Poland, Portugal, Trinidad, United Arab Emirates and United Kingdom.
- Recall number
- Z-1549-06
- Recall status
- Terminated
- CDRH record ID
- 47011
- FDA classified
- 2006-09-30
- Recall event number
- 35499