Z-1549-2024: Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium he

Recall record Z-1549-2024 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1549-2024
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Apr 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.

Source facts

Action
Siemens issued Urgent Medical Device Correction (UMDC) ACHC24-03.A.US to US customers via FedEx beginning on 03/13/2024. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 03/13/2024. New customers who purchase the Atellica CH Iron3 assay on the Atellica CH and Atellica CI analyzers will receive the UMDC/UFSN, until the product issue is resolved. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Refer to short-term mitigation steps in Appendix B while Siemens works towards transitioning customers back to the Iron_2 assay. {The UMDC/UFSN provides instructions to Atellica Solutions customers with multiple analyzers to segregate Iron3 and the impacted assays (Chol_2, LDLC, and Trig_2) onto separate systems. Atellica Solutions customers with a single Atellica analyzer instructed to stop use of Iron3 and identify alternative methods for Iron testing]. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution Siemens is actively working to ramp up the supply volume to support customer conversion to the Iron_2 assay. We anticipate being able to support an end to the mitigation within a six month timeframe. A follow up communication will be provided when Customer Actions are no longer required. These mitigations will remain in effect until the solution is provided
Initiated
2024-03-13
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Curacao, St¿Eus, Czech Republic, Denmark, Estonia, Finland, France, Germany, ¿ Greece, Hungary, India, Ireland, Israel, Italy, Latvia, Lithuania, Mexico, Netherlands, ¿¿¿¿ New Zealand, Norway, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United¿Kingdom.¿
Recall number
Z-1549-2024
Recall status
Open, Classified
CDRH record ID
206750
FDA classified
2024-04-16
Recall event number
94260

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗