Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method.
Source facts
Action
Consignees were informed via letter dated 9/15/06 to perform Folate assays in batch mode and to run controls and liquid flow cleaning upon completion of each batch run. Distributors were requested to notify their customers.
Initiated
2006-09-15
Root cause
Other
Terminated
2008-04-30
Distribution
Nationwide.