Open original FDA record ↗Recall record
- FDA identifier
- Z-1550-2008
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Jul 30, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
Source facts
- Action
- Consignees were notified via letter (Urgent: Medical Device Correction) dated 4/3/08 to identify their stocks on hand, verify package sealing, to discard any non-sealed packages, mail the Business Reply Card regardless if there are any remaining affected products.
- Initiated
- 2008-04-03
- Root cause
- Packaging process control
- Terminated
- 2009-12-17
- Distribution
- Worldwide Distribution --- including USA and countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, Ecuador, England, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Largos, Lebanon, Lithuania, Malta, Mexico, Morocco, Netherlands, Nigeria, Panama, Peru, Poland, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunis, Turkey and Venezuela.
- Recall number
- Z-1550-2008
- Recall status
- Terminated
- CDRH record ID
- 69240
- FDA classified
- 2008-07-30
- Recall event number
- 47334