Skip to content

Public FDA record

Z-1550-2022: Wolf Pak Premium Dressing Change Kit with Statlock CS/30 Ref: DC5040LF

Recall record Z-1550-2022 for Trinity Sterile, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1550-2022
Company in FDA record
Trinity Sterile, Inc.
FDA record creation date
Aug 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Kits mislabeled as latex-free, the packaging for component 94-7002 Non-Adherent Dressing contains natural rubber latex. Potential risk to to those with latex allergy

Source facts

Action
Trinity Medical issued Urgent Medical Device Correction Letter dated June 30, 2022 via email as a follow-up to the call on 6/29/22. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the affected products and lot numbers identified on the attached spreadsheet. If you have affected stock in inventory, immediately discontinue use and quarantine the products. If you have transferred any of the affected kits to a third party, please communicate this information to them. 2. Notify Trinity. Please complete the enclosed Customer Response Form as soon as possible, indicating whether you understand this medical device correction. Fax it to 410-860-2913 or email it to customerservice@trinitysterile.com. 3. Relabel the affected product. Once Trinity receives your completed Customer Response Form, Trinity will provide you with brightly colored stickers to attach to each case and kit. If you have any questions, please contact Customer Service at 410-860-5123.
Initiated
2022-06-29
Root cause
Labeling design
Terminated
2024-05-10
Distribution
US Nationwide distribution in the states of FL, PA.
Recall number
Z-1550-2022
Recall status
Terminated
CDRH record ID
195156
FDA record created
2022-08-18
Recall event number
90625

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗