Skip to content

Public FDA record

Z-1552-2014: TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-00

Recall record Z-1552-2014 for Tenex Health Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1552-2014
Company in FDA record
Tenex Health Inc
FDA classification date
May 6, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.

Source facts

Action
Tenex Health sent an Urgent Medical Device Recall notification letter dated April 15, 2014, to all consigness. The letter informed consignees of the recall, manufacturer, product name, product code, reason for recall, recall carried out to the user level, instructions to customers, contact information, label examples, and response form was provided. Consignees were instructed to immediately examine their inventory per the enclosed instructions, if they identify any of the affected product within their inventory, they should segregate the product and immediately return only MicroTips with damaged trays to the address below, using the Tenex Health, Inc. FEDEX account number and referencing the Returned Goods Authorization (RGA) number. RGA# Tenex Health, Inc. 26902 Vista Terrace Lake Forest, CA 92630 Customers were asked to complete and return the enclosed response form as soon as possible. Customers with questions should contact Tenex at (949) 454-7500, ext. 209 or email tanaka@tenexhealth.com. For questions regarding this recall call (949) 454-7500.
Initiated
2014-04-14
Root cause
Packaging
Terminated
2014-07-16
Distribution
Worldwide Distribution - USA (Nationwide) and Internationally to Panama and Singapore.
Recall number
Z-1552-2014
Recall status
Terminated
CDRH record ID
126906
FDA classified
2014-05-06
Recall event number
68074

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗