Open original FDA record ↗Recall record
- FDA identifier
- Z-1552-2020
- Company in FDA record
- QUIDEL CARDIOVASCULAR INC
- FDA record creation date
- Mar 20, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
Source facts
- Action
- On April 10, 2019 Quidel issued a recall notice to affected consignees to take the following actions: 1. Check the Quidel Triage TOX Drug Screen Control 1you have on hand to determine if you have any of the affected lot number (C3470AN). 2. If you locate the above-noted lot number, please properly destroy them to ensure they are not used. 3. If you have forwarded this lot to another facility, please forward this communication to that facility. 4. Complete the enclosed Inventory Assessment/Certificate of Destruction form and return to CustomerNotifications@quidel.com. 5. If you have any questions or require assistance, please contact Quidel Technical Support: " For US and Canada, please call 800-874-1517, option 2 " For Europe and Middle East, please call +353 {91) 412 474
- Initiated
- 2019-04-09
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2021-01-14
- Distribution
- US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL
- Recall number
- Z-1552-2020
- Recall status
- Terminated
- CDRH record ID
- 179075
- FDA record created
- 2020-03-20
- Recall event number
- 84698