Skip to content

Public FDA record

Z-1552-2020: Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per

Recall record Z-1552-2020 for QUIDEL CARDIOVASCULAR INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1552-2020
Company in FDA record
QUIDEL CARDIOVASCULAR INC
FDA record creation date
Mar 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Source facts

Action
On April 10, 2019 Quidel issued a recall notice to affected consignees to take the following actions: 1. Check the Quidel Triage TOX Drug Screen Control 1you have on hand to determine if you have any of the affected lot number (C3470AN). 2. If you locate the above-noted lot number, please properly destroy them to ensure they are not used. 3. If you have forwarded this lot to another facility, please forward this communication to that facility. 4. Complete the enclosed Inventory Assessment/Certificate of Destruction form and return to CustomerNotifications@quidel.com. 5. If you have any questions or require assistance, please contact Quidel Technical Support: " For US and Canada, please call 800-874-1517, option 2 " For Europe and Middle East, please call +353 {91) 412 474
Initiated
2019-04-09
Root cause
Software Manufacturing/Software Deployment
Terminated
2021-01-14
Distribution
US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL
Recall number
Z-1552-2020
Recall status
Terminated
CDRH record ID
179075
FDA record created
2020-03-20
Recall event number
84698

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.