Z-1553-05: Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump, catalog #2M8064, 2M8064R (refurbished), 2M806

Recall record Z-1553-05 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1553-05
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Sep 21, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.

Source facts

Action
Safety Alert letters dated 5/3/05 were sent to all Flo-Gard 6201 and 6301 Volumetric Infusion Pump customers. The letters informed the users of the potential for the pole clamp mounting screws to come loose, and recommended that they inspect the Flo-Gard pole clamp during their routine maintenance or as part of the recommended annual preventative cycle. If the pole clamp appears to be loose, the users were provided with the following steps to be followed to ensure that the pole clamp is properly secured: a) The recessed area in the housing for the clamp body must be clean. b) The three mounting screw holes in the housing must not show signs of cracking. c) The ground wire must be attached to the center screw. d) The screws must be M4x10 with captive washer. e) Loctite 425 threadlocker must be applied to each of the three screws. f) Torque the screws to 7.8 in-lb. Enclosed was a copy of the Service Bulletin issued in April 2002 with additional information about servicing the Flo-Gard pole clamps.
Initiated
2005-05-03
Root cause
Other
Terminated
2008-03-04
Distribution
Nationwide including Puerto Rico, and internationally to Canada, Taiwan, Thailand, Belgium, Switzerland, New Zealand, China, Argentina, Peru, Australia, Asia, Hong Kong and the Philippines.
Recall number
Z-1553-05
Recall status
Terminated
CDRH record ID
39473
FDA classified
2005-09-21
Recall event number
31965

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗