Open original FDA record ↗Recall record
- FDA identifier
- Z-1553-2014
- Company in FDA record
- Siemens Healthcare Diagnostics
- FDA classification date
- May 13, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
Source facts
- Action
- On February 18, 2014 Siemens Healthcare Diagnostics distributed Urgent Medical Device Correction notices to their customers via courier service (Urgent Medical Device Correction/). Upon receipt of the communication, customers are asked to follow "Actions to be Taken by the Customer". If the cover is unable to stay partially raised, immediately contact your Siemens technical support representative to arrange for a service. Complete and return the Field Correction Effectiveness Check attached to this letter within thirty (30) days by faxing to (302) 631-7597. Customers with questions should contact their Siemens technical support representative.
- Initiated
- 2014-02-18
- Root cause
- Device Design
- Terminated
- 2017-03-06
- Distribution
- Nationwide and internationally
- Recall number
- Z-1553-2014
- Recall status
- Terminated
- CDRH record ID
- 125825
- FDA classified
- 2014-05-13
- Recall event number
- 67603