Z-1553-2020: Omnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D: PDM-USA1-D001-MG-USA1 EU:

Recall record Z-1553-2020 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1553-2020
Company in FDA record
Insulet Corporation
FDA record creation date
Mar 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.

Source facts

Action
On February 14, 2020, the firm notified all affected customers of the recall via Urgent Medical Device Correction letters through email and post; the customer communication was also posted on the firm's website. Customers were informed of the product issue and provided with the following precautions: 1. Always enter a current BG into the Bolus Calculator before giving a bolus. This will replace the old BG value and ensure an accurate dose is calculated based on the current BG. 2. Total Bolus should always be the sum of the Meal Bolus and Correction Bolus. The firm will be making a software update to the next version of Omnipod DASH PDM which will provide a solution for this issue and is expected to be available starting March 2020. The firm will be voluntarily providing a replacement PDM to all existing Omnipod DASH users. The firm will contact users directly when replacements are ready to be shipped. Customers may visit www.omnipod.com for updated information on PDM replacement timing. In the meantime, the firm states that customers may continue using their current Omnipod DASH PDM provided you follow the above precautions. If you have any questions regarding the information provided in this Medical Device Correction, please contact 1-800-581-6359.
Initiated
2020-02-13
Root cause
Software design
Terminated
2021-09-14
Distribution
US Nationwide distribution. Foreign distribution to Italy, Netherlands, and United Kingdom.
Recall number
Z-1553-2020
Recall status
Terminated
CDRH record ID
179735
FDA record created
2020-03-20
Recall event number
84982

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗