Z-1553-2021: ReWalk Personal 6.0, ReWalk Personal 5.0 and ReWalk R Systems orthotically fits to the lower limbs and pa

Recall record Z-1553-2021 for Argo Medical Technologies Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1553-2021
Company in FDA record
Argo Medical Technologies Ltd
FDA record creation date
May 11, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ReWalk Robotics received two complaints (one in the US; one in the Germany), which alleged that the ReWalk device waistpack (which is the location of the Li-Ion battery packs) caught fire while being charged.

Source facts

Action
Rewalk issued Customer Communication Letter - Proper Operation and Storage of the ReWalk System Batteries to ReWalk System Owners, Companions, Distributors and ReWalk Certified Physiotherapists via certified Mail on 2/28/20. Letter states reason for recall, health risk and action to take: " General: o Do not charge the batteries while wearing the device. o If the battery usage time becomes much shorter than usual, contact your ReWalk representative. o Do not store, charge or use the system in an environment where the ambient temperatures may be higher than 113 Fahrenheit (45 Celsius) or lower than 32 Fahrenheit (0 Celsius). o Improper handling of the ReWalk Main and Aux rechargeable batteries may cause chemical leakage, heat, smoke, fire, or an explosion. " During System Charging: o When connecting the charger to the system, make sure the connector is securely screwed in place. o Do not charge the batteries continuously for more than 8 hours. If the battery is not fully charged after 8 hours - discontinue charging, stop the usage of the system and call your ReWalk Service representative. o Make sure to fully charge the system after each use. o The system is fully charged when the charger LEDs are Green. o Do not charge the system if the charger LEDs are flashing - discontinue charging, stop the usage of the system and call your ReWalk Service representative. " Charging Before System Storage: o Short term storage (7 days or less) The system must be fully charged before storage. The ReWalk system should be fully charged at least once a week. o Long term storage (more than 7 days) Fully charge the system before storage, then turn off the system, open the waistpack and disconnect the Main and Aux battery plugs (refer to the relevant steps in the User Guide). For first use after storage connect the plugs, close the waistpack and charge the system until the charger LEDs are Green. Do not charge the batteries for more than 8 hours
Initiated
2020-02-28
Root cause
Device Design
Terminated
2021-11-10
Distribution
US Nationwide distribution.
Recall number
Z-1553-2021
Recall status
Terminated
CDRH record ID
187111
FDA record created
2021-05-11
Recall event number
87818

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗