Open original FDA record ↗Recall record
- FDA identifier
- Z-1553-2021
- Company in FDA record
- Argo Medical Technologies Ltd
- FDA record creation date
- May 11, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ReWalk Robotics received two complaints (one in the US; one in the Germany), which alleged that the ReWalk device waistpack (which is the location of the Li-Ion battery packs) caught fire while being charged.
Source facts
- Action
- Rewalk issued Customer Communication Letter - Proper Operation and Storage of the ReWalk System Batteries to ReWalk System Owners, Companions, Distributors and ReWalk Certified Physiotherapists via certified Mail on 2/28/20. Letter states reason for recall, health risk and action to take: " General: o Do not charge the batteries while wearing the device. o If the battery usage time becomes much shorter than usual, contact your ReWalk representative. o Do not store, charge or use the system in an environment where the ambient temperatures may be higher than 113 Fahrenheit (45 Celsius) or lower than 32 Fahrenheit (0 Celsius). o Improper handling of the ReWalk Main and Aux rechargeable batteries may cause chemical leakage, heat, smoke, fire, or an explosion. " During System Charging: o When connecting the charger to the system, make sure the connector is securely screwed in place. o Do not charge the batteries continuously for more than 8 hours. If the battery is not fully charged after 8 hours - discontinue charging, stop the usage of the system and call your ReWalk Service representative. o Make sure to fully charge the system after each use. o The system is fully charged when the charger LEDs are Green. o Do not charge the system if the charger LEDs are flashing - discontinue charging, stop the usage of the system and call your ReWalk Service representative. " Charging Before System Storage: o Short term storage (7 days or less) The system must be fully charged before storage. The ReWalk system should be fully charged at least once a week. o Long term storage (more than 7 days) Fully charge the system before storage, then turn off the system, open the waistpack and disconnect the Main and Aux battery plugs (refer to the relevant steps in the User Guide). For first use after storage connect the plugs, close the waistpack and charge the system until the charger LEDs are Green. Do not charge the batteries for more than 8 hours
- Initiated
- 2020-02-28
- Root cause
- Device Design
- Terminated
- 2021-11-10
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1553-2021
- Recall status
- Terminated
- CDRH record ID
- 187111
- FDA record created
- 2021-05-11
- Recall event number
- 87818