Z-1553-2025: Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

Recall record Z-1553-2025 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1553-2025
Company in FDA record
Philips North America Llc
FDA classification date
Apr 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for collimator to fall as a result of incorrect installation.

Source facts

Action
Consignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).
Initiated
2025-03-10
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, Slovenia, South Africa, Spain, Thailand, United Kingdom, Vietnam.
Recall number
Z-1553-2025
Recall status
Open, Classified
CDRH record ID
212867
FDA classified
2025-04-11
Recall event number
96435

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗