Open original FDA record ↗Recall record
- FDA identifier
- Z-1554-2015
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- May 19, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Source facts
- Action
- Philips Healthcare Planned action: 1. Philips healthcare will contact customers and initiate a software update to correct the software deficiency causing the problem. 2. Philips engineers will install the software. 3. Philips engineers will perform testing to ensure the software update was effective. 4. Corrections are to be made free of charge 5. Corrective actions should be completed by August 15, 2015 For further questions call (978) 659-7892
- Initiated
- 2015-02-09
- Root cause
- Software design (manufacturing process)
- Terminated
- 2016-06-14
- Distribution
- Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
- Recall number
- Z-1554-2015
- Recall status
- Terminated
- CDRH record ID
- 135515
- FDA classified
- 2015-05-19
- Recall event number
- 70916