Open original FDA record ↗Recall record
- FDA identifier
- Z-1555-2009
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- Jun 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm received and confirmed that some lots of Cardiac Troponin I (cTnl) CalPaks (CCTNI-CR) may exhibit calibration issues that cause lower than typical signal during calibration with the current population of cTnl TestPaks (CCTNI). This issue may result in failed calibrations due to "Slope Error" messages or falsely elevated patient sample and quality control (QC) results on calibrations that
Source facts
- Action
- Siemens Healthcare Diagnostics, Inc. issued an Urgent Field Safety Notice dated April 29, 2008 to all affected customers instructing them to discontinue use of 3 CalPak lots and reduce the shelf-life for the remaining lots. For further information, contact Siemens Healthcare Diagnostics, Inc. at 1-302-631-7672.
- Initiated
- 2008-04-28
- Root cause
- Labeling False and Misleading
- Terminated
- 2009-07-14
- Distribution
- Worldwide Distribution - including United States, United Arab Emirates, Austria, Bermuda, Belize, Canada, Germany, Egypt, India, Japan, Saudi Arabia and Bengal.
- Recall number
- Z-1555-2009
- Recall status
- Terminated
- CDRH record ID
- 81883
- FDA record created
- 2015-06-25
- Recall event number
- 52007