Skip to content

Public FDA record

Z-1555-2011: Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are

Recall record Z-1555-2011 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1555-2011
Company in FDA record
Elekta, Inc.
FDA classification date
Mar 4, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent within the iCom tolerance table.

Source facts

Action
The firm, Elekta, sent an "Important Notice A302" dated August 23, 2007 titled "Update to iCom Tables for Desktop Pro when used with Third Party R&V Systems",to all affected customers. On September 28, 2007, Elekta issued a second Important Notice A304 which reiterated the same information as was in Important Notice A302. The notices described the product, problem and actions to be taken. This notice provides a workaround on how to avoid the issue. In order to avoid this issues documented, Elekta recommended that the customers set the tolerance for collimators in the iCom tolerance table to be 3.0 degrees-no less than 3.0 and to use the Linac Record, if configured, as a means of checking the final collimator position. Elekta noted that the issue will be resolved in a future release of Desktop Pro. If you any questions related to this issue, please contact your local Elekta, third party R&V system representative or Elekta Limited at +44 (0)1293 654200.
Initiated
2007-08-23
Root cause
Other
Terminated
2011-04-07
Distribution
Worldwide distribution: USA including Puerto Rico and countries including: Canada and Mexico.
Recall number
Z-1555-2011
Recall status
Terminated
CDRH record ID
97006
FDA classified
2011-03-04
Recall event number
57688

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗