Skip to content

Public FDA record

Z-1555-2015: Philips Healthcare DuraDiagnost X- Ray

Recall record Z-1555-2015 for Philips Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1555-2015
Company in FDA record
Philips Medical Systems, Inc.
FDA classification date
May 19, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).

Source facts

Action
Philips Healthcare Planned action: 1. Philips healthcare will contact customers and initiate a software update to correct the software deficiency causing the problem. 2. Philips engineers will install the software. 3. Philips engineers will perform testing to ensure the software update was effective. 4. Corrections are to be made free of charge 5. Corrective actions should be completed by August 15, 2015 For further questions call (978) 659-7892
Initiated
2015-02-09
Root cause
Software design (manufacturing process)
Terminated
2016-06-14
Distribution
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
Recall number
Z-1555-2015
Recall status
Terminated
CDRH record ID
135516
FDA classified
2015-05-19
Recall event number
70916

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗