Z-1555-2023: VITROS XT 7600 Integrated System - For use in the in vitro quantitative, semi-quantitative, and qualitati

Recall record Z-1555-2023 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1555-2023
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
May 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusable

Source facts

Action
Ortho Clinical issued Important Product Correction letter on 3/10/23 to customers with VITROS XT7600 & 5600 Integrated Systems (CL2023-061) and instructed: Until further notice, Ortho recommends loosening and then retightening all caps for MicroTip reagents or diluents immediately prior to loading them onto the system. Per the V-Docs/Maintenance Troubleshooting Guide instructions, Manual MicroTip Reagent Pack Cap Removal, Chapter 6 Troubleshooting, Remove reagent pack caps manually if they become too tight for the MICROTIP PACK OPENER to uncap. The purpose of executing this procedure is to loosen the caps without removing them, and then manually tighten the caps. For VITROS 5600: J33044 Chapter 6 Troubleshooting: Manual MicroTip Reagent Pack Cap Removal. For VITROS XT7600 Chapter 6 Troubleshooting: Manual MicroTip Reagent Pack Cap Removal " Complete the enclosed Confirmation of Receipt form no later than March 24, 2023. " Please forward this notification if the affected product was distributed outside of your facility. " Save this notification with your user documentation or post this notification by each VITROS 5600 or XT 7600 System until root cause investigation is completed. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. Contact Information Ortho Care Technical Solutions Center at 1-800-421-3311
Initiated
2023-03-10
Root cause
Under Investigation by firm
Distribution
Nationwide including Puerto Rico and Bermuda Foreign: Australia Belgium Brazil Canada Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore Spain Sweden The Netherlands United Kingdom
Recall number
Z-1555-2023
Recall status
Open, Classified
CDRH record ID
199496
FDA classified
2023-05-04
Recall event number
92055

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗