Z-1555-2025: Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (

Recall record Z-1555-2025 for GE Medical Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1555-2025
Company in FDA record
GE Medical Systems, LLC
FDA classification date
Apr 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/6/25 was sent to customers. Actions to be taken by Customer/User: You can continue to use your Venue Go and Venue Fit system after following the actions below: 1) Turn the system Off and unplug the system from the wall socket. 2) Remove the system batteries by following instructions provided in the User Manual Chapter 3. These instructions are also provided in this letter in Appendix A for Venue Go and in Appendix B for Venue Fit. 3) Check the date of manufacture of the system batteries by looking at the label on the bottom of each battery (See Figure 1). The date of manufacture is displayed as YYYY-MM for year and month of manufacture. NOTE: Venue Go has two batteries and Venue Fit has one battery 4) Calculate the age of each of the system batteries. 5) If any of the batteries are more than 2 years old: a. With the system unplugged, re-insert batteries, turn the system On and leave the system On to discharge the batteries. b. Once the battery capacity indicator shows two segments or less (See Figure 2), turn the system Off and remove the batteries per step 2. c. Dispose of batteries according to local regulations. NOTE: After battery removal, you can continue to use the system with use of AC power cord. The system is fully functional without the battery. 6) If the batteries are less than 2 years old, you may continue to use the battery until the battery reaches 2 years old, at which point, repeat instructions in step 5. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.78101@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. After the correction has been implemented, please be sure to destroy
Initiated
2025-03-06
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1555-2025
Recall status
Open, Classified
CDRH record ID
213122
FDA classified
2025-04-11
Recall event number
96538

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗