Skip to content

Public FDA record

Z-1556-2011: PrecisePLAN Treatment Planning System Intended use: To plan multiple beam radiation therapy treatments.

Recall record Z-1556-2011 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1556-2011
Company in FDA record
Elekta, Inc.
FDA classification date
Mar 4, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

When using the 'Beam Edit" function to modify a bean in an unlocked Fraction Group, there are two conditions that may cause the user to unitentionally modify a locked fraction group.

Source facts

Action
The firm, Elekta, sent an "Important Notice A312" titled "Unintentional Modification of a locked fraction group in PrecisePLAN" dated February 5, 2008 to all affected customers. The notice described the product, problem and action to be taken. The customers were instructed to use care when using the "Make Opposed" function and when changing the Fraction Group of a beam. The customers were also advised to review the final plan and any locked fraction groups to make sure the planned beam arrangement is what was intended. Note: A Field Change Order 540 073 "All PrecisePLAN 2.15 users" was issued on September 30, 2008 to all customers. If you have any questions, call 770-670-2548 or +44(0)1293 654400.
Initiated
2008-02-05
Root cause
Other
Terminated
2011-04-07
Distribution
Worldwide distribution: USA and country of: Mexico.
Recall number
Z-1556-2011
Recall status
Terminated
CDRH record ID
97014
FDA classified
2011-03-04
Recall event number
57693

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗