Skip to content

Public FDA record

Z-1557-2016: eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood

Recall record Z-1557-2016 for Philips Visicu. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1557-2016
Company in FDA record
Philips Visicu
FDA record creation date
Apr 22, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa

Source facts

Action
1. Field Safety Notice(FSN) will be sent to all customers using eCareCoordinator v.1.3 platform. The FSN describes the problem and the circumstances in which the design defects occur and the action planned by Philips to correct the problem. The FSN will also detail the functionality change in detail. 2. A software correction will be made and released. Customers will be upgraded to eCC v.1.4. The v.1.4 is scheduled to be released on or about 8 March 2016. The defect will have been corrected and tested.
Initiated
2016-03-10
Root cause
Under Investigation by firm
Terminated
2017-10-27
Distribution
US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.
Recall number
Z-1557-2016
Recall status
Terminated
CDRH record ID
144720
FDA record created
2016-04-22
Recall event number
73673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗