Open original FDA record ↗Recall record
- FDA identifier
- Z-1557-2016
- Company in FDA record
- Philips Visicu
- FDA record creation date
- Apr 22, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa
Source facts
- Action
- 1. Field Safety Notice(FSN) will be sent to all customers using eCareCoordinator v.1.3 platform. The FSN describes the problem and the circumstances in which the design defects occur and the action planned by Philips to correct the problem. The FSN will also detail the functionality change in detail. 2. A software correction will be made and released. Customers will be upgraded to eCC v.1.4. The v.1.4 is scheduled to be released on or about 8 March 2016. The defect will have been corrected and tested.
- Initiated
- 2016-03-10
- Root cause
- Under Investigation by firm
- Terminated
- 2017-10-27
- Distribution
- US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.
- Recall number
- Z-1557-2016
- Recall status
- Terminated
- CDRH record ID
- 144720
- FDA record created
- 2016-04-22
- Recall event number
- 73673