Open original FDA record ↗Recall record
- FDA identifier
- Z-1557-2020
- Company in FDA record
- Abbott Diabetes Care, Inc.
- FDA classification date
- Feb 28, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
Source facts
- Action
- On 11/14/19, Urgent Medical Device Correction notices were mailed to customers via overnight delivery. Customers were told to continue to use instructions for use regarding the cleaning of meters for use on single patients only. Customers were further asked to complete and return customer response forms. Customers with additional questions were encouraged to call customer service in 1-888-522-5226.
- Initiated
- 2019-11-14
- Root cause
- Labeling False and Misleading
- Terminated
- 2021-04-28
- Distribution
- U.S.: KY, NY, TN, MA, IL, SC, VT, FL, CA, NC, MS, NJ, TX, AL, GA, OK, CT, PA, NE, ND, MT, MO, DC, ME, LA, KS, ID, OH, MN, NM, VA, IN
- Recall number
- Z-1557-2020
- Recall status
- Terminated
- CDRH record ID
- 179577
- FDA classified
- 2020-02-28
- Recall event number
- 84866