Skip to content

Public FDA record

Z-1557-2026: HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P /

Recall record Z-1557-2026 for Covidien LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1557-2026
Company in FDA record
Covidien LLC
FDA classification date
Mar 11, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

Source facts

Action
On 01/30/2026, the firm sent via mail or email an "Urgent Medical Device Correction" letter to customers informing them that based on a customer report determined that the SpO2 No Pulse Alarm was not transmitted to the Vital Sync" system as it was not properly configured in the system software. It was further identified that two additional alarms from RespArray" bedside monitors, CO2 No Breath Detected and RESP No Breath Detected Alarms, were not properly configured in the Vital Sync" system software. This issue only affects Vital Sync" remote patient monitoring system used with a RespArray" bedside monitor. Customers are instructed to: "Notify all personnel in all care environments where RespArray" monitors are connected to Vital Sync"systems about this safety notice. "Post this notice in a prominent location and ensure awareness of the issue is maintained until the software update is installed. "If further distribution of RespArray" monitors that are connected to a Vital Sync" remote monitoring system to other facilities, promptly forward a copy of this letter to those recipients. Patient Management: For patients currently monitored by a RespArray" patient monitor and connected to the HealthCast" Vital Sync" remote monitoring system as an adjunct for patient monitoring, be advised that the SpO2 No Pulse, CO2 No Breath Detected and RESP No Breath Detected Alarm notifications from the primary RespArray" patient monitor are not transmitted to the Vital Sync" remote monitoring system. Refer to Clinical Discovery Guide for specific configurations selected by your facility. No additional action is required by treating clinicians other than recognition that the SpO2 No Pulse, CO2 No Breath Detected and RESP No Breath Detected Alarms are not available to be transmitted from the RespArray" monitor to the Vital Sync" remote monitoring system until after the updated software is installed. For question - contact Technical and Clinical Support at 800-255-6774, Option 6.
Initiated
2026-01-30
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, FL, GA, IL, IN, KY, LA, MI, MN, MS, NC, NY, OH, OK, OR, SC, TN, TX, and VA. The country of Spain.
Recall number
Z-1557-2026
Recall status
Open, Classified
CDRH record ID
218630
FDA classified
2026-03-11
Recall event number
98407

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗