The coupler on this device was manufactured without set-screws that are used to hold and center the rear assembly and the endobody. The coupler is part of the camera system and affects the ability to focus.
Source facts
Action
Only one customer had all of the defective couplers. The firm called this customer on March 27, 2008, and asked that the 28 couplers remaining in their possession be returned.
Initiated
2008-03-31
Root cause
Nonconforming Material/Component
Terminated
2008-09-17
Distribution
Product was delivered to one consignee in Iowa.
Recall number
Z-1558-2008
Recall status
Terminated
CDRH record ID
69537
FDA classified
2008-08-24
Recall event number
47463
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.