- FDA identifier
- Z-1558-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 22, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.
Source facts
- Action
- Siemens distributed a Safety Advisory Notice dated April 1, 2016, to affected customers The letter identified the affected product, problem, how to avoid potential risks, how the issue will be resolved. and actions to be taken. Customers were asked to observe the safety notice and comply with the corresponding measures until further notice.
- Initiated
- 2016-04-01
- Root cause
- Software design
- Terminated
- 2017-04-17
- Distribution
- US Nationwide Distribution : NC, NH, IL, MA, WA, PA, MD, MN, TX, NY, SC, OH, CA, KY, MT, IA
- Recall number
- Z-1558-2016
- Recall status
- Terminated
- CDRH record ID
- 145086
- FDA classified
- 2016-04-22
- Recall event number
- 73838