Open original FDA record ↗Recall record
- FDA identifier
- Z-1559-2013
- Company in FDA record
- Hitachi Medical Systems America Inc
- FDA classification date
- Jun 19, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
Source facts
- Action
- Hitachi sent a Device Correction Notice beginning on January 31,2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Software correction is targeted to start by March 1, 2013. All sites will be visited by Hitachi Service to perform corrections. All letters are recorded by tracking number and date received by the customers. For further questions please call 800-800-3106 x 3720.
- Initiated
- 2013-01-31
- Root cause
- Software design
- Terminated
- 2013-09-20
- Distribution
- Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.
- Recall number
- Z-1559-2013
- Recall status
- Terminated
- CDRH record ID
- 118892
- FDA classified
- 2013-06-19
- Recall event number
- 65438