Open original FDA record ↗Recall record
- FDA identifier
- Z-1559-2016
- Company in FDA record
- Becton Dickinson & Co.
- FDA classification date
- Apr 22, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Trypticase TM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar is contaminated with Listeria Monocytogenes. The device is a non-sterile prepared plated media.
Source facts
- Action
- BD sent an Urgent Product Recall letter dated March 2016 to all US customers via UPS or email. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the affected lots and discard any remaining product. BD will replace the discard inventory. Customers were asked to complete and fax the attached form to : Attention Regulatory Compliance, Fax: 410-316-4258 or email to BDRC2@bd.com.
- Initiated
- 2016-03-28
- Root cause
- Material/Component Contamination
- Terminated
- 2016-11-14
- Distribution
- US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, KS, LA, MA, MD, ME, MI, MN, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, VA, WA.
- Recall number
- Z-1559-2016
- Recall status
- Terminated
- CDRH record ID
- 144969
- FDA classified
- 2016-04-22
- Recall event number
- 73783