Skip to content

Public FDA record

Z-1559-2020: Intellijoint Navigation System, Software Application, IJK1 (P/N 000-0547) - Product Usage: The Intellijoi

Recall record Z-1559-2020 for Intellijoint Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1559-2020
Company in FDA record
Intellijoint Surgical, Inc.
FDA record creation date
Mar 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During total knee arthroplasty (TKA), the software may incorrectly calculate femur resection depth.

Source facts

Action
The firm notified the Intellijoint Surgical Sales Representatives that manage the affected sites. Communications were sent via email on 2/11/2020. Sales reps are responsible for verbally communicating the correction to affected users within their respective service areas. Users were to be advised of the following: (1) Until a Software update is provided, users should consult the Instructions for Use (Section 6.9.4 Lateral Femoral Condyle Definition (pg. 42) and Section 6.9.5 Medial Femoral Condyle Definition (pg. 42)) and follow the steps outlined for the acquisition of data used to calculate resection depth. (2) Resection depth should be verified by manual methods (calipers, angel wings etc.) for each case. (3) Due to the low risk and nature of the error, Intellijoint is not requesting that affected product be returned unless requested by the customer. Should the customer/user wish to return affected devices, please contact Sankavi Muralitharan (s.muralitharan@intellijointsurgical.com) or Breanne Cuddington, Senior Regulatory Affairs Associate, at b.cuddington@intellijointsurgical.com; 1-888-232-2634 ex. 2234. A software update is under development to correct for this error. The expected release date for the update is February 29, 2020, at which time Intellijoint Sales Representatives will update all affected systems with the new software by March 30, 2020.
Initiated
2020-02-11
Root cause
Software design
Terminated
2020-06-10
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY and CT. Country of Canada.
Recall number
Z-1559-2020
Recall status
Terminated
CDRH record ID
179912
FDA record created
2020-03-23
Recall event number
85016

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗