Open original FDA record ↗Recall record
- FDA identifier
- Z-1559-2020
- Company in FDA record
- Intellijoint Surgical, Inc.
- FDA record creation date
- Mar 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During total knee arthroplasty (TKA), the software may incorrectly calculate femur resection depth.
Source facts
- Action
- The firm notified the Intellijoint Surgical Sales Representatives that manage the affected sites. Communications were sent via email on 2/11/2020. Sales reps are responsible for verbally communicating the correction to affected users within their respective service areas. Users were to be advised of the following: (1) Until a Software update is provided, users should consult the Instructions for Use (Section 6.9.4 Lateral Femoral Condyle Definition (pg. 42) and Section 6.9.5 Medial Femoral Condyle Definition (pg. 42)) and follow the steps outlined for the acquisition of data used to calculate resection depth. (2) Resection depth should be verified by manual methods (calipers, angel wings etc.) for each case. (3) Due to the low risk and nature of the error, Intellijoint is not requesting that affected product be returned unless requested by the customer. Should the customer/user wish to return affected devices, please contact Sankavi Muralitharan (s.muralitharan@intellijointsurgical.com) or Breanne Cuddington, Senior Regulatory Affairs Associate, at b.cuddington@intellijointsurgical.com; 1-888-232-2634 ex. 2234. A software update is under development to correct for this error. The expected release date for the update is February 29, 2020, at which time Intellijoint Sales Representatives will update all affected systems with the new software by March 30, 2020.
- Initiated
- 2020-02-11
- Root cause
- Software design
- Terminated
- 2020-06-10
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of NY and CT. Country of Canada.
- Recall number
- Z-1559-2020
- Recall status
- Terminated
- CDRH record ID
- 179912
- FDA record created
- 2020-03-23
- Recall event number
- 85016