Open original FDA record ↗Recall record
- FDA identifier
- Z-1560-2015
- Company in FDA record
- Bard Access Systems
- FDA classification date
- Apr 30, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Bard Access Systems is conducting a field action due to the potential that an incorrect Maximum Flow rate label associated to the power injection IFU for the PowerLoc MAX Safety Infusion Set (SWIS) was attached to the outside of the unit pouch.
Source facts
- Action
- The firm, Bard, sent an "URGENT: Notification-Potential Incorrect Infusion IFU Label for PowerLoc MAX Safety Winged Infusion Set (SWIS)" letter dated March 25, 2015 to its customers. The letter described the product, problem and actions taken. The firm has corrected the cause of the issue and all future lots will be manufactured with the correct maximum flow rate of 2mL/second. Customers may return for a replacement or keep items. Please direct any return requests and questions to our coordinator at 801-522-5640.
- Initiated
- 2015-03-13
- Root cause
- Labeling False and Misleading
- Terminated
- 2015-06-24
- Distribution
- US Distribution to states of: GA and MN.
- Recall number
- Z-1560-2015
- Recall status
- Terminated
- CDRH record ID
- 135029
- FDA classified
- 2015-04-30
- Recall event number
- 70834