Skip to content

Public FDA record

Z-1560-2021: Eyesuite 9.3.1 software, LENSTAR, MODEL LS900: ac-powered slit-lamp biomicroscope.

Recall record Z-1560-2021 for Haag-Streit USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1560-2021
Company in FDA record
Haag-Streit USA Inc
FDA record creation date
May 11, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Examination data and patient name may be mixed up when printing or generating a pdf with the interface at Pacific Coast and Laser Institute (PCLI).

Source facts

Action
On 04/06/2021, Haag-Streit issued an Urgent Medical Device Correction notice to their customer via letter notifying them Eyesuite 9.3.1 software may mix up examination data and patient name when printing or generating a PDF.
Initiated
2021-04-15
Root cause
Software design
Distribution
US Nationwide distribution in the state of WA.
Recall number
Z-1560-2021
Recall status
Open, Classified
CDRH record ID
186957
FDA record created
2021-05-11
Recall event number
87766

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗