Z-1560-2022: Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator Syste

Recall record Z-1560-2022 for Getinge Usa Sales Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1560-2022
Company in FDA record
Getinge Usa Sales Inc
FDA record creation date
Aug 22, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to hypoventilation and consequently desaturation, with patient outcomes including hypoxemia and hypoxic injury, which might result in circulatory failure.

Source facts

Action
On July 20, 2022, the firm sent Urgent Medical Device Correction letters to customers. Customers were instructed to examine their inventory immediately to see if they had any of the affected serial numbers on hand. Per the company, customers may continue to use their Ventilator System in accordance to the Users manual with extra attention to the following precautions as listed in chapter 1.2 Safety Guidelines of the User s manual: o The patient must never be left unattended when connected to the ventilator system o Always make sure that a manual resuscitator is readily available The firm has updated the software platform for Servo systems to version 4.4 which contains bug fixes to address the technical alarms. If you have any questions, please contact your Maquet Critical Care AB/Getinge representative or call the Maquet Critical Care AB/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2022-07-20
Root cause
Software in the Use Environment
Distribution
Worldwide distribution. US nationwide, Poland, Spain, France, Germany, Czech Republic, Italy, Belgium, Estonia, Greece, Finland, Portugal, Netherlands, Romania, Austria, Slovakia, Lithuania, Ireland, Latvia, Norway, Sweden, Malaysia, China, Singapore, Taiwan, Iceland, Slovenia, Hungary, Bulgaria, Denmark, Croatia, United Kingdom, Russia, Hong Kong, United Arab Emirates, Canada, Switzerland, Australia, New Zealand, Serbia, Kenya, South Africa, Palestine, India, Colombia, Macau, Argentina, Mexico, Japan, Brazil, Israel, South Korea, Turkey, Peru, Ukraine, Thailand, Polynesia(French), Morocco, Saudi Arabia, Iran, Martinique (French), Costa Rica, Kuwait, Algeria, Saint Helena, Syria, Chile, Qatar, Fiji, Lebanon, Sri Lanka, Jordan, Oman, Pakistan, Ecuador, Botswana, Mozambique, Netherlands, Antilles, Bolivia, Andorra, Aruba, Faroe Islands, Egypt, Libya, Indonesia, Bahrain, Uruguay, Guadeloupe (French), Puerto Rico, Bosnia-Herzegovina, Paraguay, Nepal, Nicaragua, Vietnam, Tunisia, El Salvador, Tanzania, Papua New Guinea, Philippines, Nigeria, Myanmar, Dominican Republic, Cuba, Russian Federation, Honduras, Zimbabwe, Moldova, Republic of Bangladesh, Maldives, Angola Mauritius, Sudan, Turkmenistan, Panama, Rwanda, Brunei Darussalam, and Georgia.
Recall number
Z-1560-2022
Recall status
Open, Classified
CDRH record ID
194852
FDA record created
2022-08-22
Recall event number
90579

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗