Open original FDA record ↗Recall record
- FDA identifier
- Z-1562-2022
- Company in FDA record
- DeRoyal Industries Inc
- FDA record creation date
- Aug 22, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.
Source facts
- Action
- The recalling firm issued recall letters dated 7/15/2022 via email to the distributors and sale representative, and via overnight mail to the institution on 7/15/2022. The letter with an attached Affected Products Listing explained the reason for recall, the health risk, and provided actions for the customer to take, which included the following: Using the attached Affected Products Listing, identify the affected part numbers and lot numbers in inventory and place them in quarantine pending return. Complete the Notice of Return Form and return it to the recalling firm via FAX or email no later than 8/26/2022. If the product has been further distributed, the customer is requested to contact their customers by either forwarding the recall notification or requesting the recalling firm notify them.
- Initiated
- 2022-07-15
- Root cause
- Under Investigation by firm
- Terminated
- 2024-08-06
- Distribution
- US Nationwide distribution in the states of FL, IN, NV, RI, TN, TX, and WV.
- Recall number
- Z-1562-2022
- Recall status
- Terminated
- CDRH record ID
- 195196
- FDA record created
- 2022-08-22
- Recall event number
- 90629