Open original FDA record ↗Recall record
- FDA identifier
- Z-1563-2011
- Company in FDA record
- Hill-Rom, Inc.
- FDA classification date
- Mar 7, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Hill-Rom has become aware of a potential issue with some TotalCare and VersaCare beds equipped with the integrated P500 Microclimate Management (MCM) therapy surface, and with the VersaCare P500 MCM surface upgrade. Because of an unauthorized vendor part modification, it is possible that the MCM blower may not perform as originally designed. Given this modification, the MCM blower may experie
Source facts
- Action
- Hill-Rom sent an "IMPORTANT MEDICAL DEVICE CORRECTION" notice dated February 21, 2011, to Facility Risk Manager/Facility Administrators. The letter states that because of an unauthorized vendor part modification, it is possible that the MCM blower may not perform as originally designed. Given this modification, the MCM blower may experience the following: delayed start, slow running, or failure to operate.This issue does not pose a hazard to patients as long as normal patient monitoring is in effect in relation to managing skin temperature and moisture. Customers were instructed to continue to use the product as they normally would. The mattress system would provide an alarm in the event the MCM blower does not operate properly. A Hill-Rom representative would be contacting their facility to make arrangements for correcting the affected MCM blowers. If customers had any questions concerning this request or the procedure as outlined, they were instructed to contact Hill-Rom Technical Support at 800-445-3720. For questions regarding this recall call 812-934-7777.
- Initiated
- 2011-02-21
- Root cause
- Other
- Terminated
- 2012-08-06
- Distribution
- Worldwide Distribution - USA including AL, CA, CO, CT, DC, DE, FL, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WI, and the countries of ARUBA, AUSTRIA, CANADA, FRANCE, GERMANY, LEBANON, PORTUGAL and SAUDI ARABIA
- Recall number
- Z-1563-2011
- Recall status
- Terminated
- CDRH record ID
- 97795
- FDA classified
- 2011-03-07
- Recall event number
- 57944