Open original FDA record ↗Recall record
- FDA identifier
- Z-1563-2016
- Company in FDA record
- 3M Company - Health Care Business
- FDA record creation date
- Apr 24, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.
Source facts
- Action
- 3M sent an Urgent Medical Device Field Correction letter to customers beginning 01/22/2016. The letter identified affected product, and stated that a 3M Service representative will contact the consignee to inspect the sterilizer. Questions can be directed to 3M Health Care Service Line at 1-800-228-3957.
- Initiated
- 2016-01-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-08-25
- Distribution
- AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include: Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.
- Recall number
- Z-1563-2016
- Recall status
- Terminated
- CDRH record ID
- 143088
- FDA record created
- 2016-04-24
- Recall event number
- 73117