Z-1563-2020: VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators.

Recall record Z-1563-2020 for LivaNova USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1563-2020
Company in FDA record
LivaNova USA Inc
FDA record creation date
Mar 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.

Source facts

Action
On 02/04/20, Urgent Medical Device Correction notices were mailed to Doctors, who were informed the following: Which Patients are Potentially Impacted? Because the false positive warning message only appears if the generator is initially interrogated at 0mA, there are only certain scenarios where the issue may appear: - After initial implant, when the device is programmed on for the first time; - Following an MRI, after the device was programmed to 0mA per labeling instructions; or - Resuming therapy following previous device disablement due to an adverse event or device issue. Actions to be taken by the Physician: 1. If you obtain the low output current warning message, follow the Physician's Manual instructions to perform a System Diagnostics test after initial interrogation and programming before utilizing in-session re-interrogation. 2. During programming visits, always perform System Diagnostics testing immediately after any programming operations are performed to ensure the device is delivering programmed stimulation as intended. 2. Please complete and return the attached Customer Response Form. Please ensure that this notice is communicated to all personnel within your organization who need to be aware of it. To report a complaint or for technical support, contact the Customer Quality team at cservices@livanova.com or (866) 882-8804. Actions to be taken by Company: An update to the software is in development to eliminate this false positive low output current message. Information on availability of the software update will be shared in future communications. New users will be trained on proper use of programming system, including the appropriate sequence of use of diagnostic testing.
Initiated
2020-02-04
Root cause
Software Design Change
Distribution
U.S.: LA, ID, AZ, UT, KY, VA, TN, NY, SC, IL, IN, NC, WA, CA, OH, FL, PA, NJ, CO, TX, RI, MD, HI, OR, MA, NE, MO, OK, MS, MN, SD, VT, AR, WV, WI, MI, AL, KS, DE, NM, GA, ND, AK, MT, ME, NV, DC, MO, WY, IA, CT. Canada, and Israel.
Recall number
Z-1563-2020
Recall status
Open, Classified
CDRH record ID
179188
FDA record created
2020-03-24
Recall event number
84723

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗