Z-1564-2014: Siemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 tests) Cat No. 06367974 SMN 10310308 (10

Recall record Z-1564-2014 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1564-2014
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA classification date
May 7, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Serum controls may go out of range low with the ADVIA Centaur Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223

Source facts

Action
Siemens issued An Urgent Medical Device Recall (UMDR) to US customers on March 25, 2014. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on March 24, 2014. These notices inform the customer of reported controls failures with Folate ReadyPack kit lots ending in 223. In addition the notices inform customers of a shift when moving from Folate ReadyPack kit lots ending in 219 and kit lots ending in 222, 224, 225, 226 and 227. Customers were instructed to discontinue use of ADVIA Centaur Systems Folate kit lots ending in 219 and 223. ADVIA Centaur systems Folate kit lots ending in 222, 224, 25, 226 and 227 are performing within specification and may continue to be used for folate interpretation. The notices should be reviewed with the facility's laboratory or medical director. Fax-back forms will be used to confirm customers received the notification.
Initiated
2014-03-25
Root cause
Nonconforming Material/Component
Terminated
2015-04-30
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria Belgium Bulgaria Cyprus Czech Republic Denmark Estonia Finland Finland France Fren.Polynesia Germany Great Britain Greece Guadeloupe Hungary Iceland Ireland Israel Italy Latvia Lithuania Luxembourg Malta Martinique Netherlands Norway Poland Portugal Qatar Reunion Romania Russian Fed. Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir.
Recall number
Z-1564-2014
Recall status
Terminated
CDRH record ID
126468
FDA classified
2014-05-07
Recall event number
67839

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗