Open original FDA record ↗Recall record
- FDA identifier
- Z-1565-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Aug 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the remote console joystick where a fatal accident occurred on a non-GE Remote R&F System. .
Source facts
- Action
- Two letters "GE Product Safety Notification" dated 12/18/07 and 3/7/08, addressed to Hospital Administrators, Manager of Radiology and Radiologists, were sent to each consignee. Each letter addressed a separate safety issue. Letters described the affected product, safety instructions, corrective action and contact information. The joystick can become stuck causing unintentional motion of the tube. Field Modification Instructions (FMI 10802/10803) have been issued to the field to install additional warning labels on the patient table, install a protector for the console joystick and inspect and correct the clearance between the console joystick and its mounting location. If your facility has a Precision RXi or Precision e system manufactured before March 2007, a GE Healthcare Field Engineer will schedule a field visit to install the added joystick protection. This upgrade will be performed at no charge. If you have any questions/concerns regarding this issue, please contact your local GE Healthcare sales or service representative.
- Initiated
- 2007-12-18
- Root cause
- Device Design
- Terminated
- 2012-08-09
- Distribution
- (USA) AK CO IL LA MA MI MS NC NJ NY PA SC TN UT VA WA WI (OUSA) BR CA MX AU C3 KR PH TW AE AT BE BF BG CM CY CZ DE ES FI FR GB GE GR GY HR HU IT LB LV MT NA NG PL PT RO RU SA SD SE SI SK SY TN TR UA YU ZA
- Recall number
- Z-1565-2008
- Recall status
- Terminated
- CDRH record ID
- 70060
- FDA classified
- 2008-08-24
- Recall event number
- 46807