It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field:
Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal.
Pin: The pin used to guide the Lock Shaft Inserter into the ma
Source facts
Action
Spine Frontier sent an Urgent Notification dated December 23, 2011, consignees. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached acknowledgement form. For questions call 866-914-7717
Initiated
2011-12-23
Root cause
Device Design
Terminated
2014-05-20
Distribution
US Nationwide - OR, FL, MA, TX, CO, OK, IL, and MO
Recall number
Z-1565-2014
Recall status
Terminated
CDRH record ID
112689
FDA classified
2014-05-07
Recall event number
63145