Skip to content

Public FDA record

Z-1565-2022: Nobel Biocare N1 TiUltra TCC NP 3.5x13mm

Recall record Z-1565-2022 for Nobel Biocare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1565-2022
Company in FDA record
Nobel Biocare
FDA classification date
Aug 23, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.

Source facts

Action
On 07/25/2022, Nobel Biocare AB sent an "URGENT MEDICAL DEVICE RECALL" Letter via UPS to customers informing them that there may be a burr present on the co-packed OsseoShaper 1 around the waist of the device. The hazard is primarily identified as a risk of the patient swallowing or inhaling the burr if it comes loose from the OsseoShaper 1 during the clinical procedure. Customer are instructed to: 1. Inspect their stock and quarantine affected devices. 2. Complete attached Customer Acknowledgment Form, even if they do not have any affected stock, and return it to Nobel Biocare, via email to us.cs.complaints@nobelbiocare.com or fax 714-282-5073 within 5 days of receipt of this notice. 3. Return all affected stock on hand to Nobel Biocare using the shipping label attached to this notice. 4. Ensure relevant staff members are informed of this recall. If they have supplied or transferred any potentially affected product to another facility or organization, let that facility know of the recall by providing a copy of this notice. Nobel Biocare has implemented preventive actions and will replace devices subject to this recall free of charge. For questions - contact local customer support representative at 800-322-5001 ext. 1527.
Initiated
2022-07-25
Root cause
Process change control
Distribution
U.S Nationwide.: AZ, CA, CT, FL, HI, NC, NY, OR, PA, TX, UT, and WI O.U.S.: Austria, Belgium, Canada, Germany, Spain, France, United Kingdom, Hungary, Iceland, Italy, Netherlands, Norway, Portugal, and Sweden
Recall number
Z-1565-2022
Recall status
Open, Classified
CDRH record ID
195403
FDA classified
2022-08-23
Recall event number
90679

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗