Open original FDA record ↗Recall record
- FDA identifier
- Z-1565-2026
- Company in FDA record
- Abbott Diagnostics Scarborough, Inc.
- FDA classification date
- Mar 13, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was confirmed that the impacted lot has a higher occurrence of invalid rates when compared to the product Instructions for Use.
Source facts
- Action
- On or about February 9, 2026 Product Removal Notice letters were sent to customers and distributors. Actions to be taken: " Quarantine and discontinue use of any remaining inventory of the impacted lot. " Count how many unused kit boxes (24 tests / 1 kit box) of remaining impacted inventory require replacement. " Complete and return the Customer Acknowledgment Form (Form must be fully completed, signed and returned to receive replacement product). " Complete the Customer Request for Replacement Product Form. " Customer Service will arrange replacement product and return of your product. " Retain this letter for your records. forward this notice to those you have further distributed or transferred product to.
- Initiated
- 2026-02-09
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1565-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218604
- FDA classified
- 2026-03-13
- Recall event number
- 98400