Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.
Source facts
Action
The firm sent an "URGENT VOLUNTARY DEVICE REMOVAL" letter dated March 18, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Initiated
2013-03-18
Root cause
Process design
Terminated
2013-07-17
Distribution
Worldwide Distribution-US Distribution in New Jersey and the countries of Germany, Switzerland, Italy, Norway, The Netherlands, Spain, Portugal, Ireland, and Austria.
Recall number
Z-1566-2013
Recall status
Terminated
CDRH record ID
118594
FDA classified
2013-06-21
Recall event number
65343