Open original FDA record ↗Recall record
- FDA identifier
- Z-1567-2008
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Aug 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Batteries mislabeled - batteries labeled as 2.4 amp/hour batteries, however are 1.0 amp/hour batteries.
Source facts
- Action
- On February 27, 2008, the firm sent a "URGENT MEDICAL DEVICE - ACCESSORY RECALL" letter via certified return receipt to their consignees. The letter states the batteries are fully functional 1.0 amp/hour batteries mislabeled as 2.4 amp/hour batteries and that the batteries will provide half the monitoring time and half the number of defibrillation shock as of a 2.4 amp/hour LIFEPAK NiCd battery. Enclosed with the letter is a confirmation sheet for the consignee to indicate the number of recalled batteries they have and fax back to the firm. The firm will send out replacement batteries and the consignee is to then send back the mislabeled batteries. If there are any questions, contact the technical support group at 1-800-442-1142 - option 5 or visit www.physio-control-notices.come/fastpak.
- Initiated
- 2008-02-27
- Root cause
- Labeling mix-ups
- Terminated
- 2011-03-07
- Distribution
- Batteries were distributed nationwide and to Canada, Netherlands, Asia Pacific and Latin America.
- Recall number
- Z-1567-2008
- Recall status
- Terminated
- CDRH record ID
- 68503
- FDA classified
- 2008-08-24
- Recall event number
- 46921