Open original FDA record ↗Recall record
- FDA identifier
- Z-1567-2020
- Company in FDA record
- LeMaitre Vascular, Inc.
- FDA classification date
- Apr 1, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.
Source facts
- Action
- On March 23, 2020, the firm distributed "URGENT: MEDICAL DEVICE RECALL" letters to affected customers. Customers were notified of the risk of the catheter failing to deflate, which may ultimately lead to tip separation. Customers were asked to do the following: 1. Check your inventory and immediately quarantine any recalled devices. 2. Complete the form at the end of this letter. Please note that you must return the form even if you have no devices in inventory. 3. Scan the reply form and send it to recalls@lemaitre.com. 4. If you have recalled devices at your facility, LeMaitre Vascular, Inc. will send you instructions for returning the devices. 5. When the recalled devices have been received by LeMaitre Vascular, we will arrange for replacement material. 6. If you have transferred devices to another facility, please forward a copy of this recall letter to them. Customers with questions may contact the Senior Quality Engineer, at rlerer@lemaitre.com or 781-221-2266. ***Updated 8/21/20*** Firm added Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246 to the recall. These lots were previously on hold in a warehouse in Germany, but in July they were mistakenly distributed to customers in Europe.
- Initiated
- 2020-03-23
- Root cause
- Process control
- Distribution
- Worldwide distribution: US (Nationwide distribution) and foreign countries of: Australia, Austria, Belgium, Bulgaria, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Georgia, Hungary, Ireland, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, and Vietnam.
- Recall number
- Z-1567-2020
- Recall status
- Open, Classified
- CDRH record ID
- 180253
- FDA classified
- 2020-04-01
- Recall event number
- 85193