Skip to content

Public FDA record

Z-1567-2020: LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. Th

Recall record Z-1567-2020 for LeMaitre Vascular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1567-2020
Company in FDA record
LeMaitre Vascular, Inc.
FDA classification date
Apr 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.

Source facts

Action
On March 23, 2020, the firm distributed "URGENT: MEDICAL DEVICE RECALL" letters to affected customers. Customers were notified of the risk of the catheter failing to deflate, which may ultimately lead to tip separation. Customers were asked to do the following: 1. Check your inventory and immediately quarantine any recalled devices. 2. Complete the form at the end of this letter. Please note that you must return the form even if you have no devices in inventory. 3. Scan the reply form and send it to recalls@lemaitre.com. 4. If you have recalled devices at your facility, LeMaitre Vascular, Inc. will send you instructions for returning the devices. 5. When the recalled devices have been received by LeMaitre Vascular, we will arrange for replacement material. 6. If you have transferred devices to another facility, please forward a copy of this recall letter to them. Customers with questions may contact the Senior Quality Engineer, at rlerer@lemaitre.com or 781-221-2266. ***Updated 8/21/20*** Firm added Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246 to the recall. These lots were previously on hold in a warehouse in Germany, but in July they were mistakenly distributed to customers in Europe.
Initiated
2020-03-23
Root cause
Process control
Distribution
Worldwide distribution: US (Nationwide distribution) and foreign countries of: Australia, Austria, Belgium, Bulgaria, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Georgia, Hungary, Ireland, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, and Vietnam.
Recall number
Z-1567-2020
Recall status
Open, Classified
CDRH record ID
180253
FDA classified
2020-04-01
Recall event number
85193

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.