Skip to content

Public FDA record

Z-1567-2022: Stryker SWANSON Flexspan Finger Joint Implant W/O Grommets Silicone, Size: 00, REF 4700020, Sterile.

Recall record Z-1567-2022 for Wright Medical Technology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1567-2022
Company in FDA record
Wright Medical Technology, Inc.
FDA record creation date
Aug 23, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An incorrect sizing label was applied to the carton packaging.

Source facts

Action
The recalling firm issued letters dated 7/27/2022 to Health Care Professionals, Operators of Medical Devices, and Distributors. Distributors received the letter via email. The medical facility received the letter via FedEx. The letter informed the consignee of the product affected, product description, product issue, potential risks, and the actions needed by customers and distributors. The actions included (1) Immediately check their internal inventory to locate the product and remove it from their point of use and isolate/quarantine it to prevent accidental usage; (2) Return the enclosed business reply form by email to confirm receipt of the notification and document product segregation. The form is to be returned by 8/10/2022; (3) Return all affected product in their possession. Hospitals are instructed to contact their local sales office or Stryker sales representative directly for product returns and inventory questions. Branches/Hub are to return the affected devices to the address specified in the letter; (4) Ensure that personnel within the facility are aware of this communication and all required actions are completed; and (5) If product has been further distributed, notify the applicable parties at once regarding the recall.
Initiated
2022-07-27
Root cause
Error in labeling
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, GA, NY, PA, and TN. The countries of the Australia, France, the Netherlands, Sweden, and the United Kingdom.
Recall number
Z-1567-2022
Recall status
Open, Classified
CDRH record ID
195354
FDA record created
2022-08-23
Recall event number
90660

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗