Open original FDA record ↗Recall record
- FDA identifier
- Z-1568-2008
- Company in FDA record
- 3M Company / Medical Division
- FDA classification date
- Aug 8, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A production specification discrepancy may cause the indicators to show an inaccurate result, which could incorrectly lead customers to conclude the sterilization cycle was adequate.
Source facts
- Action
- Customers and Distributors were each sent a "Voluntary Product Recall" letter dated 3/17/08 on 3/19/08. The letter states that use of the affected product should immediately be discontinued, stocks should be checked for any remaining inventory, and reprocess any unused packs using the product not from the affected lot or from an alternative supplier. The letter requested consignees to notify 3M whether the facility has affected product or not by sending the "Product Return Form". Replacements will be provided to those with affected products remaining in their stock.
- Initiated
- 2008-03-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-03-01
- Distribution
- Nationwide Distribution --- including states of AL, AR, AZ, CA, CO, CT,DC, DE, FL, GA,HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS,MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV.
- Recall number
- Z-1568-2008
- Recall status
- Terminated
- CDRH record ID
- 68615
- FDA classified
- 2008-08-08
- Recall event number
- 46950