Open original FDA record ↗Recall record
- FDA identifier
- Z-1568-2016
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA record creation date
- Apr 26, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for unexpected motion of the Xi Stapler jaws on the da Vinci Xi System with p5 software relating to a combination of the p5 software and the surgeon quickly transitioning from the clamp to the fire pedal during use.
Source facts
- Action
- Intuitive Surgical sent an Urgent Product Safety notification to customers via e-mail on March 25, 2016. Customers were advised to STOP USE on the EndoWrist Stapler 45 and 30 instruments on da Vinci XI systems with p5 software. Product letters followed and were mailed on March 31, 2016 by Fed Ex. Urgent Medical Device letters were sent via Fed Ex on March 31, 2016 (dated April 1, 2016). These letters reiterated what was explained in the previous e-mail and provide more information of the unexpected motion issue. In addition to the stop use request, letters requested that all personnel be fully informed of the issue. Further information or support concerning this issue, please contact your Clinical Sales Representative or contact ISI Customer Service at the number: North America: 800-876-1310 Option 3 (6 am to 5pm PST)
- Initiated
- 2016-03-25
- Root cause
- Software design
- Terminated
- 2016-07-01
- Distribution
- US Nationwide Distribution in the states of: WI, FL, OH, TX, CA, AZ, IL, MD, MN, NY, OK, HI, MO, IA, KS, PA, and AL.
- Recall number
- Z-1568-2016
- Recall status
- Terminated
- CDRH record ID
- 144835
- FDA record created
- 2016-04-26
- Recall event number
- 73736