Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.
Source facts
Action
Terumo sent beginning May 30, 2013, an URGENT MEDICAL DEVICE RECALL NOTIFICATION were sent to all customers with instructions to notify all customers of the recall requesting all affected units to be quarantined until the heatsink units are replaced Confirm receipt by filling out the response form containing communication by faxing or emailing the attached Affiliate Response Form to the fax number or email address indicated on the form.
Questions should be directed to Terumo CVS customer service in the U.S.:
Telephone: +1-734-663-4145; Fax: +1-734-741-6149
Initiated
2013-05-30
Root cause
Employee error
Terminated
2013-09-17
Distribution
Worldwide Distribution including the countries of UAE (United Arab Emirates) HONG KONG and THAILAND. Not distributed in USA
Recall number
Z-1569-2013
Recall status
Terminated
CDRH record ID
116407
FDA classified
2013-06-21
Recall event number
64530